Izixhobo zomculoIimbonakalo:
1.Uxinzelelo olubi loxinzelelo kwikhabhinethi yebhayoloji kwindawo ekhuselekileyo yokusebenza, ukuqinisekisa ukhuseleko lwabaqhubi;
2. Igumbi elisebenzayo elinoxinzelelo oluphezulu, inqanaba elimbini lokucoca okuphezulu, i-100% yokukhutshwa okukhuselekileyo;
3. Yamkela iisampulu ze-Anderson ezine-channel ezimbini zenqanaba lesithandathu;
I-4.Impompo ye-peristaltic eyakhelwe ngaphakathi, ubukhulu bokuhamba kwepompo ye-peristaltic kulungelelaniswa;
5. Ijeneretha ye-aerosol ekhethekileyo ye-microbial, ubungakanani bebhaktheriya bolwelo lokutshiza bunokulungelelaniswa, umphumo we-atomization ulungile;
6. Ulawulo lwesikrini sombala omkhulu weshishini, ukusebenza lula;
7. I-interface ye-USB, inkxaso yokudluliselwa kwedatha;
8. I-RS232 / i-Modbus interface esemgangathweni, inokufikelela kulawulo lwangaphandle.
9. Ikhabhinethi yokhuseleko ixhotyiswe ngokukhanyisa kwe-LED, ukujonga lula;
10. Isibane sokubulala iintsholongwane se-UV esakhelwe ngaphakathi;
11. Front switch uhlobo olutywiniweyo ucango lweglasi, kulula ukusebenza kunye nokuqwalasela;
12. Nge-software esebenzayo ye-SJBF-AS, unokulawula nokusetyenzwa kwedatha ngekhompyutha,
13. Inkqubo yolawulo yolwazi lwebhubhoratri engenamthungo.
Iiparamitha zobuGcisa:
Iiparamitha eziphambili | Umda weParameter | Isigqibo | Ukuchaneka |
Ukuhamba kwesampulu | 28.3 L/min | 0.1 L/min | ±2% |
Ukutshiza | 8 ~ 10 L/min | 0.1 L/min | ±5% |
Ukuhamba kwepompo yePeristaltic | 0.006~3 mL/min | 0.001 mL/min | ±2% |
Uxinzelelo phambi kwesampulu ye-flowmeter | -20 ~ 0 kPa | 0.01 kPa | ±2% |
Uxinzelelo lwe-flowmeter yangaphambili | 0 ~ 300 kPa | 0.1kPa | ±2% |
Uxinzelelo olubi lwegumbi le-aerosol | -90 ~ -120 Pa | 0.1Pa | ±1% |
Ubushushu bokusebenza | 0~50 ℃ | ||
Uxinzelelo olubi lweKhabhinethi | > 120Pa | ||
Umthamo wokugcina idatha | Umthamo onokunyuka | ||
Ukusebenza okuphezulu kokucoca umoya | ≥99.995%@0.3μm,≥99.9995%@0.12μm | ||
Iitshaneli ezimbini 6-inqanaba le-Anderson sampler Ubungakanani bamasuntswana obanjiweyo | Ⅰ>7μm, Ⅱ4.7~7μm, Ⅲ3.3~4.7μm, Ⅳ2.1~3.3μm Ⅴ1.1~2.1μm Ⅵ0.6~1.1μm | ||
Lilonke inani lamasuntswana olawulo lwesampulu esemgangathweni | 2200±500 cfu | ||
Idayamitha ephakathi yobunzima bejenereyitha yeaerosol | Umndilili wedayamitha yesuntswana (3.0±0.3 µm), ukutenxa okusemgangathweni kwejometri ≤1.5 | ||
Inqanaba elithandathu lesampulu ye-Anderson ibamba ubungakanani besuntswana | Ⅰ>7 µm; Ⅱ(4.7~7 µm); Ⅲ(3.3~4.7 µm); Ⅳ(2.1~3.3 µm); Ⅴ(1.1~2.1 µm); Ⅵ(0.6~1.1 µm) | ||
Iimpawu zegumbi le-aerosol | L 600 x Ф85 x D 3mm | ||
Ukuhamba komoya wekhabhathi yoxinzelelo olubi | >5m3/min | ||
Ubungakanani benjini engundoqo | Ngaphakathi: 1000 * 600 * 690mm Ngaphandle: 1470 * 790 * 2100mm | ||
Ingxolo yomsebenzi | < 65db | ||
Ukusebenza kombane | AC220±10%,50Hz,1KW |