IzixhoboIimbonakalo:
1. Indawo yokusebenza ekhuselekileyo yekhabhinethi yebhayoloji yoxinzelelo olubi, ukuqinisekisa ukhuseleko lwabaqhubi;
2. Igumbi lokusebenza elinoxinzelelo oluphezulu, isihluzo esisebenza kakuhle esinamanqanaba amabini, ukukhutshwa okukhuselekileyo kwe-100%;
3. Sebenzisa isampulu ye-Anderson enamanqanaba amathandathu enamachaneli amabini;
4.Impompo ye-peristaltic eyakhelwe ngaphakathi, ubungakanani bokuhamba kwempompo ye-peristaltic buyalungiseka;
5. Ijenereyitha ye-aerosol ekhethekileyo yeentsholongwane, ubungakanani bokuhamba kwe-bacterial liquid spray bunokulungiswa, isiphumo se-atomization silungile;
6. Ulawulo lwesikrini sokuchukumisa esinombala omkhulu wezemveliso, kulula ukusebenza;
7. Ujongano lwe-USB, inkxaso yokudluliselwa kwedatha;
8. Ujongano oluqhelekileyo lwe-RS232/Modbus, lunokufikelela kulawulo lwangaphandle.
9. Ikhabhathi yokhuseleko ixhotyiswe ngezibane ze-LED, kulula ukuyijonga;
10. Isibane sokubulala iintsholongwane se-UV esakhelwe ngaphakathi;
11. Ucango lweglasi oluvaliweyo olufana nolungaphambili, kulula ukulusebenzisa nokuyijonga;
12. Ngesoftware esebenzayo ye-SJBF-AS, unokulawula kunye nokucubungula idatha ngekhompyutha,
13. Inkqubo yolawulo lolwazi lwelebhu yokubeka imithwalo ngaphandle komthungo.
Iiparameter zobugcisa:
| Iiparameter eziphambili | ububanzi bepharamitha | Isisombululo | Ukuchaneka |
| Ukuhamba kwesampulu | 28.3 L/ngomzuzu | 0.1 L/ngomzuzu | ±2% |
| Ukuhamba kwespray | 8 ~ 10 L/ngomzuzu | 0.1 L/ngomzuzu | ±5% |
| Ukuhamba kwempompo yePeristaltic | 0.006~3 mL/ngomzuzu | 0.001 mL/ngomzuzu | ±2% |
| Uxinzelelo ngaphambi kokuba kuthathwe isampuli ye-flowmeter | -20 ~ 0 kPa | 0.01 kPa | ±2% |
| I-spray flowmeter yoxinzelelo lwangaphambili | 0 ~ 300 kPa | 0.1kPa | ±2% |
| Uxinzelelo olubi lwegumbi le-aerosol | -90 ~ -120 Pa | 0.1Pa | ±1% |
| Ubushushu bokusebenza | 0~50 ℃ | ||
| Uxinzelelo olubi lweKhabhinethi | > 120Pa | ||
| Umthamo wokugcina idatha | Umthamo onokukwazi ukwandiswa | ||
| Ukusebenza kakuhle kwesihluzi somoya | ≥99.995%@0.3μm,≥99.9995%@0.12μm | ||
| Isampulu ye-Anderson enamachaneli amathandathu aneenqanaba ezimbini Ubungakanani besuntswana elibanjiweyo | Ⅰ>7μm, Ⅱ4.7~7μm, Ⅲ3.3~4.7μm, Ⅳ2.1~3.3μm , Ⅴ1.1~2.1μm , Ⅵ0.6~1.1μm | ||
| Inani lilonke lamasuntswana eesampuli zolawulo lomgangatho ezilungileyo | 2200±500 cfu | ||
| Ububanzi obuphakathi bobunzima bejenereyitha ye-aerosol | Ububanzi obuqhelekileyo be-particle (3.0±0.3 µm), ukuphambuka okuqhelekileyo kwejometri ≤1.5 | ||
| Isampulu ye-Anderson enamanqanaba amathandathu ibamba ubungakanani beesuntswana | Ⅰ>7 µm; Ⅱ(4.7~7 µm); Ⅲ(3.3~4.7 µm); Ⅳ(2.1~3.3 µm); Ⅴ(1.1~2.1 µm); Ⅵ(0.6~1.1 µm) | ||
| Iinkcukacha zegumbi le-Aerosol | Ubukhulu obuyi-L 600 x Ф85 x D 3mm | ||
| Ukuhamba komoya kwikhabhinethi yoxinzelelo olubi | >5m3/umzuzu | ||
| Ubungakanani benjini ephambili | Ngaphakathi:1000*600*690mm Ngaphandle:1470*790*2100mm | ||
| Ingxolo yomsebenzi | < 65db | ||
| Umbane osebenzayo | I-AC220±10%,50Hz,1KW | ||