I-YYT1000 I-Bacterial Filtration Efficiency Detector ayihambelani kuphela neemfuno zobuchwepheshe beemaski zotyando lwezonyango YY0469-2011 isihlomelo B kwindlela yovavanyo lobuchule bokucoca (BFE) ibhaktiriya kuqala b. 1.1.1 isixhobo sovavanyo, kodwa sikwahambelana noluntu lwaseMelika lwezinto zokuvavanya i-ASTMF2100, ASTMF2101, iimfuno zemigangatho ye-EN14683 yaseYurophu, uphuculo olutsha lwenziwa ngesiseko soku, kunye nesampulu ephindwe kabini yokuthelekisa ipneumatic ngaxeshanye, ukuphucula ukuchaneka kwesampulu, ilungele ukulinganisa kunye nokuvavanya amasebe, amaziko ophando lwezenzululwazi, abavelisi bemaski kunye namanye amasebe afanelekileyo ukuvavanya ukusebenza kwe-mask bacterial filtration.
YY0469
Q/0212 ZRB003-2015
ISO/DIS 22611
I-ASTMF2101
EN14683
1.Inkqubo yokuvavanya uxinzelelo olubi ukuqinisekisa ukhuseleko lwabaqhubi;
I-2.Yakhiwe-kwimpompo ye-peristaltic kwikhabhathi yoxinzelelo olubi, A, B ezimbini - indlela, i-Andersen ezintandathu;
3.Isantya sokuhamba kwe-spray sasebusuku sinokumiselwa kwaye umphumo we-atomization ulungile.
4.Ulawulo lwe-microcomputer olunesantya esiphezulu; i-intshi eyi-10.4 ye-intshi ye-intshi ye-intshi yokukhanya okuphezulu kombala wesikrini sokuchukumisa;
I-5.USB interface, inkxaso Ukugcinwa kwedatha yediski;
6.Ikhabhinethi izibane eziqaqambileyo eziqaqambileyo;Iswitshi yokukhusela ukuvuza okwakhelwe ngaphakathi ukukhusela ukhuseleko lwabasebenzisi;
7.Uluhlu lwangaphakathi lwekhabhinethi lwenziwe ngensimbi engenasici, uluhlu lwangaphandle lufafazwa ngeplastiki kunye ne-cold-rolled, kunye nemigangatho yangaphakathi nangaphandle i-insulated kunye ne-flame retardant.
8.Umnyango weglasi wokutshintsha ngaphambili ulungele ukuba umntu ozama ukujonga kwaye asebenze.
9.Isibiyeli esinokususwa, ukuphakama kwesibiyeli esihlengahlengiswayo; Ukuxhasa kunye nokuhambisa iicaster ezineenjongo ezimbini.
Inkcazo engundoqo | Uluhlu lokuchaza | Isigqibo | Ukuchaneka |
UkuHamba kweSampulu yeNdlela | 28.3L/min | 0.1L/min | Ngaphakathi kwe ± 2.5% |
B UkuHamba kweSampulu yeNdlela | 28.3L/min | 0.1L/min | Ngaphakathi kwe ± 2.5% |
Ukutshiza Ukugeleza | (8~10)L/min | 0.1L/min | Ngaphakathi kwe ± 2.5% |
Ukuhamba kweMpompo yePeristaltic | (0.006~3.0)mL/min | 0.001ml/min | Ngaphakathi kwe ± 2.5% |
Uxinzelelo lwangaphambili lweNdlela Isampulu yeFlowmeter | (-20~0)kPa | 0.01kPa | Ngaphakathi kwe ± 2.5% |
Uxinzelelo lwangaphambili lweNdlela ye-B Isampulu Flowmete | (-20~0)kPa | 0.01kPa | Ngaphakathi kwe ± 2.5% |
Uxinzelelo lwangaphambili lwe Tshaza i-Flowmeter | (0~300)kPa | 0.01kPa | Ngaphakathi kwe ± 2.5% |
Ubushushu obugciniwe | (-40~99)℃ | 0.1℃ | Ngaphakathi kwe ± 2.5% |
Uxinzelelo olubi lweAerosol Igumbi | (-90~-120)Pa | 0.1Pa | Ngaphakathi ±2.0% |
Uxinzelelo olubi lweKhabhinethi | -50 ~ -200pa
| ||
Ukukwazi ukugcina iDatha | > 100000 iiseti
| ||
I-Vortex Mixer Test Tube Inkcazo kunye noMyinge | Φ16×150mm ityhubhu yovavanyo,8pcs | ||
Ukusebenza okuphezulu kweNcwadana yoMoya Iinkcazelo Ngeempawu zesihluzi | Ukusebenza kakuhle kwesihluzi≥99.99% kumasuntswana> 0.3μm | ||
Idayamitha ephakathi yobunzima bejenereyitha yeaerosol | I-avareji yedayamitha: (3.0±0.3)μm; Umgangatho weJiyometri ubunzima bejenereyitha ukutenxa≤1.5
| ||
Iindlela ezimbini 6-lever Andersen izixhobo zeSampulu | Inqanaba1>7μm; Inqanaba 2: (4.7~7)μm; Inqanaba 3: (3.3 ~ 4.7)μm; Inqanaba 4: (2.1~3.3)μm; Inqanaba 5: (1.1~2.1)μm; Inqanaba 6: (0.6~1.1)μm | ||
Ubungakanani beGumbi leAerosol
| 600×85×3mm (Ubude×Ububanzi×Ukutyeba)
| ||
Lilonke inani lomgangatho olungileyo lawula amasuntswana esampula | (2200±500)cfu
| ||
I-Negative Chamber Flow Speed | ≥5m3/min
| ||
Ubungakanani boxinzelelo olubi Umnyango Wegumbi(W×D) | 1000×730mm | ||
Ubungakanani bomatshini ongundoqo (W×D×H) | 1180×650×1300mm
| ||
Ubungakanani beNkxaso Isakhelo(W×D×H) | 1180 × 650 × 600mm, Ukuphakama ngaphakathi kwe-10cm kulungelelaniswa | ||
Ingxolo yoMhloli | <65dB(A) | ||
Ubunzima | 150kg | ||
Ukunikezwa Amandla | AC220V±10%, 50Hz | ||
Ukutya umbane | <1500W |